Thursday, April 8, 2010

DART 100S Low Cost easy to operate. Interface to your LC/MS


NO solvent, NO columns, NO need to worry about sample solubility, just dip a melting point into your reaction mix and do DART!

Reference

Direct Analysis in Real Time for Reaction
Monitoring in Drug Discovery
Chris Petucci, Jason Diffendal, David Kaufman, Belew Mekonnen, Gene Terefenko,‡ and Brian Musselman*
Discovery Analytical Chemistry and Medicinal Chemistry, Wyeth Research, Collegeville, Pennsylvania, and IonSense, Inc.*, Saugus, Massachusetts

Abstract

Direct analysis in real time (DART) is a novel ionization technique that provides for the rapid ionization of small molecules under ambient conditions. In this study, several commercially available drugs as well as actual compounds from drug discovery research were examined by LC/UV/ESI-MS and DART interfaced to a quadruple mass spectrometer. For most compounds, the molecular ions observed by ESI-MS were observed by DART/MS. DART/MS was also studied as a means to quickly monitor synthetic organic reactions and to obtain nearly instantaneous molecular weight confirmations of final products in drug discovery.

For simple, synthetic organic transformations, the trends in the intensities of the mass spectral signals for the reactant and product obtained by DART/ MS scaled closely with those of the diode array or the total ion chromatogram obtained by LC/UV/ESI-MS. In summary, DART is a new tool that complements electrospray ionization for the rapid ionization and subsequent mass spectral analysis of compounds in rug discovery.

No comments: